Nov 10, 2023
Evaluation of transcutaneous near
Scientific Reports volume 13,
Scientific Reports volume 13, Article number: 4537 (2023) Cite this article
612 Accesses
3 Altmetric
Metrics details
Sudden cardiac arrest (SCA) is a leading cause of mortality worldwide. The SCA-to-resuscitation interval is a key determinant of patient outcomes, highlighting the clinical need for reliable and timely detection of SCA. Near-infrared spectroscopy (NIRS), a non-invasive optical technique, may have utility for this application. We investigated transcutaneous NIRS as a method to detect pentobarbital-induced changes during cardiac arrest in eight Yucatan miniature pigs. NIRS measurements during cardiac arrest were compared to invasively acquired carotid blood pressure and partial oxygen pressure (PO2) of spinal cord tissues. We observed statistically significant decreases in mean arterial pressure (MAP) 64.68 mmHg ± 13.08, p < 0.0001), spinal cord PO2 (38.16 mmHg ± 20.04, p = 0.0028), and NIRS-derived tissue oxygen saturation (TSI%) (14.50% ± 3.80, p < 0.0001) from baseline to 5 min after pentobarbital administration. Euthanasia-to-first change in hemodynamics for MAP and TSI (%) were similar [MAP (10.43 ± 4.73 s) vs TSI (%) (12.04 ± 1.85 s), p = 0.3714]. No significant difference was detected between NIRS and blood pressure-derived pulse rates during baseline periods (p > 0.99) and following pentobarbital administration (p = 0.97). Transcutaneous NIRS demonstrated the potential to identify rapid hemodynamic changes due to cardiac arrest in periods similar to invasive indices. We conclude that transcutaneous NIRS monitoring may present a novel, non-invasive approach for SCA detection, which warrants further investigation.
Sudden cardiac arrest (SCA) is a leading cause of mortality worldwide, accounting for 15–20% of all natural deaths across the USA and Western Europe1,2. SCA is defined as the sudden cessation of effective ventricular contractions, leading to inadequate cardiac output and hemodynamic collapse3. The loss of systemic circulation is fatal if not immediately detected and treated by initiating high-quality resuscitative measures4. Resuscitation is generally most successful when performed within five minutes of circulatory collapse5,6,7. Current estimates suggest that only 25% of SCAs are bystander-witnessed. Hence, in most cases, no witnesses are available to provide life-saving treatment or contact emergency medical services (EMS), prolonging the time it takes to initiate resuscitative measures8. Consequently, when EMS professionals arrive on the scene, they elect not to attempt resuscitation in approximately 50% of cases, as life-saving measures are deemed futile at this stage8. In attempts to reduce time-to-resuscitation intervals, there is a need to develop cost-effective, wearable systems that can noninvasively monitor parameters related to cardiac activity. This has the potential to reduce average time-to-resuscitation intervals through rapid identification of SCA and automatic notification to medical dispatchers. The use of such technology is of particular interest in high-risk populations, including patients with a prior history of cardiac arrest, congenital heart disease or cardiomyopathy2.
Near-infrared spectroscopy (NIRS) is a noninvasive optical biosensing technique that can monitor real-time changes in tissue oxygenation and hemodynamics transcutaneously9,10,11,12,13,14. NIRS technology is based on similar physical principles to photoplethysmography (PPG), a technology that has found routine application in clinical practice with the emergence of pulse oximetry15,16,17,18. PPG sensors employ two light sources (commonly one red/green and one infrared light emitter) to evaluate arterial blood oxygenation by detecting local blood volume changes arising from systolic cardiac activity19,20. These pulsatile changes in light absorption are detected by a photodetector and converted into an estimate of the arterial oxygen saturation (SpO2) based on the optical path length of the medium and the Beer–Lambert Law19,21. One drawback of PPG sensors is that they require pulsatile blood flow to give a reliable estimate of oxygen saturation, which may limit their use in settings where the pulse is weak or absent, such as cardiac arrest22,23,24.
Using principles similar to pulse oximetry, continuous-wave NIRS employs near-infrared light (between 650 to 1000 nm) to penetrate tissue and measure the amount of light absorbed by tissue chromophores (mostly hemoglobin and water)25. With the use of multiple source-detector pairs in a spatially-resolved configuration, NIRS can monitor changes in tissue oxygen delivery and consumption by measuring the following parameters: oxyhemoglobin (O2Hb), deoxyhemoglobin (HHb), total hemoglobin (THb = O2Hb + HHb; a measure of local blood volume changes), and the hemoglobin difference (Hbdiff = O2Hb-HHb, an indirect measure of oxygen utilization)]26. Furthermore, it is possible to derive an absolute estimate of the local tissue oxygen saturation26,27,28 (tissue saturation index or TSI; expressed in %) that reflects contributions from arterial (25%), capillary (5%), and venous (70%) compartments alike29. Importantly, with sampling rates of 10 Hz and above, transcutaneous NIRS sensors can detect and monitor the effect of cardiac contractile activity on tissue hemodynamics similar to PPG-based monitoring; however, unlike PPG-based technologies, pulsatile blood flow is not a prerequisite to obtaining an estimate of tissue oxygenation, which means that NIRS may have better utility in SCA detection and monitoring12.
The study aimed to evaluate whether transcutaneous NIRS could detect changes in pulse and tissue oxygenation during pentobarbital-induced cardiac arrest in our porcine model. After IV pentobarbital administration, changes in NIRS-derived tissue saturation index [TSI (%), absolute value] and oxygenated hemoglobin (O2Hb, relative value) were continuously monitored. We compared these results to gold-standard invasive blood pressure monitoring and an invasive spinal cord oxygenation monitor. Such information is essential for evaluating NIRS-based monitoring as a candidate for out-of-hospital SCA detection.
This investigation was done in coordination with a study to evaluate tissue oxygenation in spinal cord contusion-compression injuries30. All animal protocols and procedures performed in this study were approved by the Animal Care Committee of the University of British Columbia (UBC) and were compliant with the policies of the Canadian Council of Animal Care, the U.S. Army Medical Research and Material Command (USAMRMC), and Animal Care and Use Review Office (ACURO). The UBC Center for Comparative Medicine established the anesthesia and analgesia protocols. The study complied with ARRIVE guidelines.
Data from eight female Yucatan miniature pigs (weighing 24–31 kg) were analyzed as part of other NIRS studies investigating spinal cord hemodynamics after spinal cord injury (SCI). Animals were prepared for surgery, intubated and anesthetized as previously described31,32. Animals were pre-medicated with intramuscular Telazol (4–6 mg/kg), xylazine (1 mg/kg), and atropine (0.02–0.04 mg/kg). For anesthesia induction, propofol (2 mg/kg) or isoflurane (2–3% in O2) was used before the animals underwent endotracheal intubation. Propofol (8 mg/kg/h), fentanyl (12 µg/kg/h), ketamine (11 mg/kg/h) and midazolam (0.1–0.5 mg/kg/h) were used for anesthesia maintenance through a continuous rate infusion (CRI) and adjusted at the discretion of a veterinarian. Animals were mechanically ventilated with a ventilator rate of 10–12 breaths/min and a tidal volume of 6–10 mL/kg with 1.4 L (70%) nitrogen and 0.6 L (30%) of oxygen (Veterinary Anesthesia Ventilator model 2002, Hallowell EMC, Pittsfield, MA). Standard monitoring of the animals was performed throughout the procedure, including monitoring of blood pressure, end-tidal carbon dioxide, heart rate, and oxygen saturation.
The carotid artery and jugular vein were exposed by blunt dissection. The carotid artery was catheterized (20 Gauge Arterial Catheterization Set FA-04018; Arrow International, Reading, PA, USA) to monitor invasive blood pressure at a frequency of either 10 Hz or 100 Hz (depending on the experimental protocol). Mean arterial pressure (MAP) was calculated as a weighted average from the arterial blood pressure (1/3 systolic + 2/3 diastolic). These data were captured using LabChart Pro software (AD Instruments, Colorado Springs, Colorado, US). The jugular vein was catheterized (6–8 French Multi-Lumen Central Venous Catheterization Set CE-12703; Arrow International) to obtain central venous access for the infusion of IV medication during surgery. Subsequently, the animal was flipped, and a dorsal laminectomy was performed, with the animal being instrumented bilaterally with pedicle screws. In 5/8 animals, a spinal cord contusion-compression injury was induced at the level of T10 as previously described30.
Data on spinal cord partial pressure of oxygen (PO2) was collected invasively using a set of intraparenchymal sensors (NX-BF/OF/E; Oxford Optronics, Oxford, UK), which enabled us to compare changes in oxygenation measured superficially (by NIRS) to those measured from inside the spinal cord parenchyma. Each probe contains three sensors at the tip (temperature, blood flow and PO2; tip diameter 450 μm). The sensor measures tissue PO2 employing a fluorescence quenching and fiberoptic technique33, recorded at a sampling rate of 1 Hz. Depending on the experimental protocol, one or two intraparenchymal (IP) probes were inserted into the spinal cord parenchyma (Fig. 1c), as described in Cheung et al.30. The probes were inserted with the probe tips situated within the spinal cord white matter at a depth of approximately 3–4 mm under the dura. Probe placement was verified via ultrasound imaging (L14-5/38, 38 mm linear array probe, Ultrasonix RP; BK Ultrasound, Richmond, BC, Canada and VisualSonics Mx400, 30MgHz, Toronto, Canada). The probes were connected to OxyLab/OxyFlow channel monitors (Oxford Optronics, Oxford, UK), and data were streamed into LabChart Pro software (AD Instruments, Colorado Springs, Colorado, US).
Representative experimental set-up. (a) Placement of NIRS sensor (PortaMon, Artinis Medical Systems, The Netherlands) on the left hindlimb muscle of a Yucatan minipig. (b) The underside of the NIRS sensor consists of one photodiode (light receiving element) and three LEDs (each transmitting light at two wavelengths; ~ 765 nm and ~ 850 nm). The interoptode distances from the photodiode to each LEDs are as follows: 30 mm, 35 mm and 40 mm. The three LEDs emit a total optical power of 0.88 mW, 1.78 mW and 3.67 mW, respectively. (c) Intra-operative placement of spinal cord intraparenchymal (IP) probes to measure local changes in spinal cord PO2 in response to euthanasia. IP probes were placed as part of other NIRS studies. IP probes intraparenchymal probes, PO2 partial pressure of oxygen, NIRS near-infrared spectroscopy.
A transcutaneous NIRS device was used during each surgery (Fig. 1a,b). A wearable compact NIRS sensor (PortaMon/PortaLite, Artinis Medical Systems, the Netherlands) was placed on the biceps femoris muscle belly of the animal's hindlimb. The NIRS sensor is a dual-wavelength (765 and 850 nm), continuous-wave NIRS system containing three pairs of LEDs in a spatially resolved spectroscopy (SRS) configuration and a single photodiode. Based on the SRS configuration and the modified Beer–Lambert law27, we obtain information on the following parameters: oxyhemoglobin (O2Hb, μM/L), deoxyhemoglobin (HHb, μM/L), total hemoglobin (THb = O2Hb + HHb, μM/L), and the hemoglobin difference (Hbdiff = O2Hb-HHb, μM/L)]. Furthermore, we can estimate the local tissue oxygen saturation (tissue saturation index or TSI, expressed in %) based on Eq. (1), following correction for tissue scattering. The reduced scattering coefficient was adjusted to reflect that of the human musculature (no adjustments were made to account for potential differences in human vs porcine tissue).
NIRS data were collected using OxySoft version 3.0.95 software (Artinis Medical Systems, the Netherlands) at a sampling frequency of 10 Hz. Each NIRS sensor was secured using double-sided adhesive tape to avoid inducing pressure on the underlying tissue (Fig. 1a). Subsequently, the NIRS sensor was covered by a sterile surgical drape for isolation from ambient light during data collection.
The animals were euthanized with IV pentobarbital sodium administered using central jugular venous access at a standard dose of 120 mg/kg over 5 s. Pentobarbital is a short-acting barbiturate and sedative-hypnotic with a long history of use in veterinary practice for anesthesia and euthanasia34. The primary mode of action is causing GABAA-mediated CNS dysfunction, critically depressing medullary centres in the brainstem, leading to rapid onset apnea34,35, and cardiovascular collapse with features of bradycardia36,37, refractory hypotension35,37 and reduced myocardial contractility37. The start time of pentobarbital administration was marked in LabChart. During pentobarbital administration, pigs continued to receive mechanical ventilation and oxygenation. Thus, the mechanism of cardiac arrest was not secondary to the pentobarbital-induced apnea (i.e., respiratory arrest, which can lead to secondary cardiac arrest) but rather primary cardiac collapse. Upon cessation of a cardiac rhythm on transesophageal ECG, ventilation was discontinued, and death was confirmed by a registered veterinary technician upon auscultation for breathing and heart sounds and checking for a palpable peripheral pulse (ear or tail). During this time, all monitoring remained switched on, and NIRS, invasive blood pressure, and spinal cord PO2 signals continued to be captured for up to 5 min post-euthanasia.
All data were resampled at 10 Hz for subsequent analysis. In all animals, NIRS and LabChart data were aligned offline by the presence of artifact patterns in the signals. Such patterns result from various motion artifacts and can be used to identify common events in the time series measurements. Data were assessed for normal distribution using a D’Agostino Pearson omnibus normality test. The time of euthanasia was identified from an event marker inserted into LabChart during surgery. The analysis window was identified as 10 min before the event marker was inserted and then 5 min after the event had been added (15 min total). Relative NIRS parameters (O2Hb, HHb, THb and Hbdiff) were z-score normalized to the first 5 min of this window and are expressed in standard deviation units (s.d.u.). We examined mean arterial pressure (MAP), NIRS TSI (%), and spinal cord PO2 values in the last 10 s of the recording after euthanasia and compared them to average pre-euthanasia baseline values (collected over 60 s) using a paired Student's t-test (level of significance p < 0.05).
Comparisons between the time of pentobarbital infusion to the first detectable change in NIRS TSI (%), spinal cord PO2 vs invasive MAP signals were calculated using a paired Student's t-test. We defined the "first detectable change" in the descending signal as the time from euthanasia marker insertion in LabChart to the first drop of the signal two standard deviations below its baseline (the baseline value was defined as the mean value during the last-minute pre-pentobarbital infusion; see Fig. 4a).
To obtain pulse rate and amplitude from carotid arterial blood pressure (ABP) catheter measurements and NIRS O2Hb signal dynamics (selected due to prominent cardiac components in their respective time series), short-time Fourier transform (STFT) analyses (fs = 10 Hz; FFT length of 528 samples; window overlap of 24%) were carried out on z-score normalized signals. Three separate 60-s analysis windows were selected. These represent (1) baseline signals, (2) the 60-s interval immediately following pentobarbital infusion and (3) post-euthanasia signals. Pulse rates (Hz) and pulse amplitudes (dB) were derived from these STFT windows (displayed as mean ± 95% confidence intervals). Group differences (ABP vs O2HB-derived pulse amplitude and pulse rate) were analyzed using a two-tailed, unpaired Student's t-test or a repeated measure ANOVA (level of significance p < 0.05) for multiple group comparisons.
All analysis was carried out using GraphPad Prism 9.0.0 software (GraphPad Software, La Jolla, CA, USA), MATLAB (version R2021a, Natick, Massachusetts, USA) or Python libraries (Python Software Foundation). Data are presented as mean ± standard deviation (SD) (or ± standard error of the mean (SEM) when otherwise specified).
All animal protocols and procedures used for this study were in accordance with the policies of the Canadian Council on Animal Care and were reviewed and approved by the Institutional Animal Care and Use Committee (IACUC) of the University of British Columbia (protocol No. A20-0217 and A16-0311, approved Oct 26, 2020, and Feb 17, 2017, respectively), and the Animal Care and Use Review Office (ACURO) of the U.S. Army and the Naval Information Warfare Centre [Award No. N6600120-2-4046, approved Dec-21-2020 (NIWC, USAMRMC Protocol No. NIWC-7783.e002), and Award No. W81XWH-16-1-0602, approved May-16-2017, respectively].
Compared to baseline, carotid MAP decreased by 64.68 mmHg ± 13.08 in the 5 min post-pentobarbital administration (n = 7; Student's t-test, p < 0.0001; Table 1, Figs. 2a, 3a). The mean decrease in spinal cord PO2 was 38.16 mmHg ± 20.04 (n = 6; Student's t-test, p = 0.0028; Table 1, Figs. 2b, 3b). Similarly, we observed a statistically significant drop in the TSI (%) (n = 8; Student's t-test, p < 0.0001; Table 1, Figs. 2c,e, 3c), which dropped on average by 14.50% ± 3.80 within the first 5 min. Relative NIRS measures showed (Fig. 2d,f): a decrease in oxygenated hemoglobin (O2Hb; − 84.33 s.d.u. ± 38.65); a concomitant increase in the deoxygenated hemoglobin portion (HHb; 87.66 s.d.u. ± 65.62); and a decrease in the parameter of the hemoglobin difference (Hbdiff; − 127.23 s.d.u. ± 67.75) in all animals.
Transcutaneous NIRS, carotid blood pressure and spinal cord PO2 responses to pentobarbital administration. Averaged traces from (a) MAP (n = 7 animals, ± SEM), (b) spinal cord PO2, (n = 6, ± SEM), (c) TSI (%) (n = 8, ± SEM), and (d) relative changes in NIRS parameters of O2Hb, HHb, THb and Hbdiff (n = 8 ± SEM). The 5-min timepoint (dashed line) denotes the time of pentobarbital administration. (e,f) Delta changes to euthanasia relative to a 60-s pre-euthanasia baseline. TSI (%) tissue saturation index, Hbdiff hemoglobin difference, O2Hb oxygenated hemoglobin, HHb deoxygenated hemoglobin, THb total hemoglobin, MAP mean arterial pressure, PO2 partial pressure of oxygen, BSL baseline, EUT euthanasia.
Pre- and post-euthanasia comparison of carotid MAP, spinal cord PO2 and TSI (%) measures. Box plots (IQR + min and max values) of pre-euthanasia (BSL; averaged over 60 s pre-pentobarbital infusion) and post-euthanasia values (EUT; averaged over the last 10 s of the recording) in (a) MAP (mmHg) (n = 7, p < 0.0001) (b) spinal cord PO2 (mmHg) (n = 6, p = 0.0028, and (c) TSI (%) (n = 8, p < 0.0001). Group comparisons were analyzed using a paired student's t-test (level of significance p < 0.05). TSI (%) tissue saturation index, MAP mean arterial pressure, PO2 partial pressure of oxygen, BSL baseline, EUT euthanasia.
The time to first change in MAP was observed at 10.43 ± 4.73 s (Fig. 4b). NIRS-derived TSI (%) dropped at 12.04 ± 1.85 s (Fig. 4b). In 5/6 animals (as information on spinal cord PO2 was unavailable in 2/8 animals), spinal cord PO2 measures were the slowest to respond, taking 25.02 ± 9.237 s to register a change in spinal cord PO2 two standard deviations below the baseline (Fig. 4b). The difference in detection time between MAP and NIRS-derived TSI (%) (n = 7, paired Student's t-test, p = 0.3714, level of significance p < 0.05) was not significant.
Euthanasia-to-first change in hemodynamics for MAP, spinal cord PO2 and NIRS-derived TSI (%) and O2Hb. (a) Representative figure of a NIRS TSI (%) trace pre- and immediately following pentobarbital infusion. The first detected drop in the signal was defined as the first value two standard deviations below baseline (established over the last 60 s prior to pentobarbital administration). (b) Time (sec) to the first change in carotid MAP, spinal cord PO2 and hindlimb TSI (%) signals in response to euthanasia ± SD. MAP vs TSI (%) (n = 7; p = 0.3714), and MAP vs spinal cord PO2 (n = 6, p = 0.0006) were analyzed using paired student's t-tests (level of significance p < 0.05); TSI (%) tissue saturation index, MAP mean arterial pressure, PO2 partial pressure of oxygen.
In all animals, O2Hb and ABP-calculated pulse rates were comparable (see Fig. 5 for representative STFT analysis), and no statistically significant difference was detected between O2Hb and ABP-derived pulse rates within pre-euthanasia (p > 0.99), euthanasia (p = 0.97), and post-euthanasia analysis windows (n = 8, Student's t-test, level of significance p < 0.05; Table 2, Fig. 6d–f). In 6/8 animals, we noted a transient increase in pulse rate upon pentobarbital infusion (see Supplementary Fig. S1–S3, S5–7). In 2/8 animals, the pulse rate started to decline immediately after euthanasia induction (see Supplementary Figs. S4 and S8).
Representative short-time Fourier transform analyses. STFT colormaps of (a) carotid ABP and (c) NIRS O2Hb signals over a 15-min window. Pentobarbital infusion commenced at 600 s. (b,d) Three analysis windows were selected for subsequent analysis, denoting different stages of euthanasia. O2Hb oxygenated hemoglobin, ABP arterial blood pressure, BSL baseline, EUT euthanasia, STFT short time Fourier transform (for STFT colourmaps from individual animals, see Supplementary Figs. 1–8).
Comparison of ABP and NIRS-derived pulse rate and amplitude changes in response to pentobarbital administration. Mean ABP and O2Hb-derived pulse amplitudes (dB) (n = 8) during (a) baseline (p = 0.0014) and (b) euthanasia (p = 0.0009) (Student's t-test, level of significance p < 0.05). (c) ABP and O2Hb pulse amplitudes (dB) at BSL vs EUT (Repeated measure ANOVA, adjusted p-values: ABP BSL vs EUT p = 0.0047 and O2Hb BSL vs EUT p = 0.0132; level of significance p < 0.05). Mean ABP and O2Hb-derived pulse rates (Hz) (n = 8) during (d) baseline (p > 0.99), (e) euthanasia (p = 0.97) and (f) post-euthanasia (unable to calculate p). Group comparisons were analyzed using a student's t-test (level of significance p < 0.05). ns not significant, O2Hb oxygenated hemoglobin, ABP arterial blood pressure, BSL baseline, EUT euthanasia.
Pulse amplitudes (dB) were stronger in ABP compared to NIRS O2Hb signals (Fig. 6a,b). This difference was significant during baseline (n = 8, Student's t-test, p = 0.0014, level of significance p < 0.05; Fig. 6a) and following pentobarbital administration (n = 8, Student's t-test, p = 0.0009, level of significance p < 0.05; Fig. 6b). Lastly, we saw statistically significant decreases within ABP and NIRS-derived O2Hb pulse amplitudes in response to pentobarbital administration compared to baseline (n = 8, repeated measure ANOVA, adjusted p-values: ABP BSL vs EUT p = 0.0047 and O2Hb BSL vs EUT p = 0.0132; level of significance p < 0.05; Fig. 6c).
Using a porcine model of primary cardiac arrest, we examined changes in NIRS measurements during cardiac arrest and compared them to invasive measurements of spinal cord oxygenation and blood pressure monitoring. While there have been several reports using cerebral NIRS monitoring to assess return to spontaneous circulation (ROSC) and quality of resuscitation post-SCA38,39,40,41,42,43,44,45, the use of transcutaneous NIRS monitoring in the setting of cardiac arrest detection has not been explored and presents a potentially novel field of research. Our results demonstrated a quick and clear change in NIRS signals during and after cardiac arrest, similar to invasive hemodynamic measurements. Overall, these findings demonstrate the utility of muscle-NIRS-based monitoring for SCA detection; however, further studies are needed to establish its efficacy.
In this study, we used pentobarbital sodium to induce euthanasia in our porcine model. Pentobarbital depresses medullary centres in the brainstem, causing rapid death by respiratory depression34,35 and cardiovascular collapse35,36,37. There are two main models of cardiac arrest: (1) primary cardiac arrest with a direct insult to cardiac tissue, causing cessation of effective cardiac output and resultant whole-body hypoxia; and (2) primary respiratory arrest, causing whole-body hypoxia (including that of cardiac tissue), resulting in secondary cardiac arrest. Although phenobarbital causes respiratory depression, in this study, mechanical ventilation and oxygenation were continued after pentobarbital administration (preventing hypoxia), consistent with primary cardiac arrest. Further research is required to investigate physiological changes with respiratory arrest leading to secondary cardiac arrest, as is more frequently encountered in infants, children, and adults with opioid-related cardiac arrest or other respiratory etiologies46. Since NIRS measures tissue oxygenation, it may have high utility in detecting respiratory arrest before cardiac arrest.
We observed considerable inter-animal variability in pre-euthanasia measures of spinal cord PO2, which may reflect variations in the IP probes’ placement and small sampling area (reported to be approximately 0.5–1 mm3,47). These limitations result in spinal cord PO2 readings that may be highly sensitive to differences in tissue composition and local microvasculature. Similarly, baseline MAP measures displayed notable variability, which is not surprising given the wide range of physiological blood pressures. TSI (%) pre-euthanasia was comparable across animals, indicating that regional oxygen saturation is preferentially maintained within a specific range in the animal's hindlimb in the anesthetized animal.
While MAP and spinal cord PO2 dropped to levels below those observed for living tissue within the first 5 min post-euthanasia induction, we observed a smaller TSI (%) decrease (14.50% ± 3.80). In one instance, TSI (%) dropped from 76.08% to 62.63% (animal #8, Table 1). Due to a comparatively high TSI (%) at baseline, this decrease associated with complete cessation of cardiac function in this animal could be within the physiological TSI (%) range of another animal (see animals #1–2, Table 1). This suggests that a particular TSI (%) value may not be sufficient to detect sudden cardiac events but instead gains diagnostic value if interpreted in the context of NIRS-derived hemodynamic trends, pulse rate changes, and patient-specific baseline TSI (%) measures. Importantly, TSI (%) does not reflect arterial oxygen saturation (i.e. the amount of oxygen in the arterial stream available to tissues and organs), but rather is an index of oxygenated to total hemoglobin at a microvascular tissue level48 composed of both arterial and venous components and thus most likely includes "inaccessible" oxygen trapped within venous compartments. Furthermore, it is crucial to consider that other local and systemic conditions could lead to similar decreases in the TSI (%). This should be emphasized when placed in the broader context of discussing the specificity of using NIRS-derived changes in the setting of SCA detection. Our group has recently shown in our porcine model that controlled periods of hypoxia to a SpO2 target of 70% lead to a TSI% decrease of 10.5 ± 1% as recorded by the PortaMon NIRS sensor from the animal's hindlimb49. While clinically, such severe episodes of hypoxia are unlikely to be encountered on a routine basis; other investigators have reported decreases in NIRS-derived tissue saturation indices in various clinical settings that lead to tissue hypoperfusion, including vascular disease50 and shock51,52,53.
We further observed that the rate of change for both MAP and spinal cord PO2 was highest within the first minute post-pentobarbital infusion once a significant decrease in the signal below baseline was detected. This is consistent with previous reports from a canine study where pentobarbital caused MAP to decrease to undetectable levels at a mean time of 52.6 s54. The rate of TSI (%) change was slower and showed a slightly decreasing trend at 5 min post-euthanasia. MAP reflects the mean blood pressure over a single cardiac cycle and is thus highly susceptible to the loss of effective ventricular contractions and decreased cardiac output. On the other hand, NIRS measures the "effect" of this loss of contractile activity on the underlying tissue, which at the time of cardiac arrest is still metabolically active55,56. Without systemic circulation, this continued oxygen exchange may underlie the steady decline in NIRS-derived muscle oxygen saturation over a prolonged period.
Notably, the comparison between the time from pentobarbital administration to the first detectable drop in NIRS-derived TSI (%) from a pre-defined baseline demonstrates the high sensitivity of NIRS to detect systemic changes in tissue oxygenation non-invasively. In contrast, the first detectable drop in spinal PO2 occurred significantly slower. Spinal cord PO2 represents the local partial pressure of oxygen (i.e., "dissolved" oxygen) in the tissue and thus provides a readout of oxygen availability at a cellular level. As oxygen is initially released from hemoglobin binding sites to meet the metabolic demand of the tissue, local spinal cord PO2 will remain stable until the metabolic demand of the tissue surpasses the oxygen-releasing capacity of hemoglobin. However, other factors should also be considered, such as potential delays in IP probe response times and low sampling rates.
We also compared mean NIRS O2Hb- and carotid ABP-derived pulse amplitudes at three different time points (pre-, during- and post-euthanasia). In all animals, ABP-derived pulse amplitudes were stronger than those derived from NIRS, both before and immediately following intravenous pentobarbital infusion. This is unsurprising as both signals are derived from different sources. Invasive blood pressure monitoring is achieved by monitoring cyclical pressure changes from one of the major vessels bifurcating from the aorta. NIRS, on the other hand, measures pulse changes indirectly by capturing the effect of pulsatile blood flow on underlying tissue dynamics at a site distal to the heart. As previously discussed, NIRS will capture changes occurring on a microvascular level, capturing contributions from arteriolar, capillary, and venular blood alike. Importantly, venules and capillaries do not contribute significantly to pressure waveform propagation while still making up a significant portion of the NIRS signal. This will contribute to the reduced pulse amplitude observed in the NIRS signal compared to the ABP. Of note is that in six out of eight animals, we observed the presence of cardiac activity after death had been clinically confirmed, consistent with previous reports that such signals may persist for 5.5–16 min after pentobarbital infusion54. Notably, both ABP- and NIRS-derived mean pulse amplitudes decreased during euthanasia (compared to baseline), indicating a decrease in cardiac inotropic activity captured by both systems. Real-time NIRS-based pulse amplitude monitoring presents a potentially new avenue of how cardiac arrest may be identified, encouraging further investigation.
We also observed a strong agreement between NIRS and carotid ABP-derived pulse rates during the three distinct analysis windows, as revealed by short-time Fourier transform analyses. Notably, in six out of eight animals, we observed transient increases in heart rate during and immediately following the infusion of pentobarbital, consistent with previous findings54. This suggests that, while overall NIRS-derived pulse amplitudes are weaker at this more distal body site, the NIRS system captures changes in pentobarbital-induced pulse rate pre-and during cardiac dysfunction at a high sensitivity when assessed against invasive blood pressure monitoring. Based on these results, we conclude that transcutaneous NIRS may be used in the setting to provide similar information. We could not perform this assessment on IP-derived parameters, as the low sampling frequency of the combined OxyLite/OxyFlow system prevented us from detecting pulsatile blood flow changes occurring on a beat-to-beat basis.
Several other technologies have been examined for non-invasively detecting out-of-hospital SCA. These include using wearable ECG-based monitoring/defibrillation systems that have shown promise in reducing mortality in high-risk patients when worn following myocardial infarction57 and left ventricular dysfunction58. Furthermore, photoplethysmography (PPG) smartwatch/bracelet-based heart rate monitoring systems are available59,60,61,62,63. However, as PPG sensors rely on pulsatile blood flow, they become less reliable or may be unable to return a measurement during periods of systemic hypotension, hypothermia or poor peripheral circulation24,64,65. As such, PPG monitoring has been predominately investigated for its ability to capture changes in heart rate and heart rate variability as a way to monitor return to spontaneous circulation (ROSC) during cardiopulmonary resuscitation59,66 and the long-term monitoring and screening of cardiac arrhythmia as previously reviewed62. The ability of NIRS to capture the cessation of cardiac pulsations while monitoring tissue oxygenation changes (even in the absence of a pulse waveform) suggests that NIRS could be a helpful tool to detect cardiac arrest in an out-of-hospital setting and track its progression until ROSC is resumed. This would further facilitate investigation into a potential relationship between NIRS-derived oxygenation changes during cardiac arrest progression and clinical prognosis. Lastly, smart devices have been used to capture agonal breathing patterns to detect SCA67. Agonal breathing is a brainstem reflex resulting from severe hypoxia. However, agonal breathing only occurs in 55% of witnessed cardiac arrest cases68. The resulting low sensitivity constraints widespread use of such monitoring in the setting of SCA detection.
Overall, our results suggest that NIRS may have utility for the rapid detection of cardiac arrest. Recent technical advancements in electro-optic components and NIRS technology enable the development of compact, wearable, low-power, sensitive, cheap, and flexible NIRS sensors. Integrating miniaturized NIRS sensors with other relevant wearable biosensors may form novel multi-modal sensing approaches for extensive monitoring of vital signs and prediction and rapid diagnosis of critical cardiac dysfunctions.
There are several limitations to what we report. First, while there are many different etiologies and physiological variations within human SCA, our pentobarbital-induced porcine model of euthanasia SCA may differ from the human experience. This is particularly relevant, as the time from pentobarbital infusion to confirmed death is unknown, which limits our ability to extrapolate findings to a more clinically relevant scenario. Secondly, since physiological data were captured with different recording software and were subsequently aligned offline by the presence of artifact patterns in the signal, they are susceptible to minor errors in alignment. However, given the unique nature of artifact patterns in the signals, we expect minimal misalignment within data sets, if any. Similarly, we expect some inter-animal variability from discrepancies in the exact timing of euthanasia marker insertion denoting IV pentobarbital administration. Lastly, it is essential to note that no adjustments to scattering coefficients were made to account for potential human vs porcine differences within the calculation of the tissue saturation index, which is established for human use. Without making model adjustments, we postulate that the following inter-species differences could affect the calculation and should be further explored in subsequent studies: (1) skin pigmentation, (2) skin texture, (3) skin thickness and (4) subcutaneous adipose tissue thickness.
Transcutaneous NIRS monitoring may present a novel, feasible, non-invasive SCA detection approach that warrants further investigation. The ability of transcutaneous NIRS to simultaneously capture the cessation of cardiac pulsatile activity while monitoring changes in tissue oxygenation may result in the development of novel, wearable optical sensors for the early detection of SCA and its subsequent tracking. This may improve the management of unwitnessed SCAs and accelerate access to life-saving interventions.
The datasets generated during and/or analyzed during the current study are available from the corresponding author upon reasonable request.
Srinivasan, N. T. & Schilling, R. J. Sudden cardiac death and arrhythmias. Arrhythm. Electrophysiol. Rev. 7, 111 (2018).
Article PubMed PubMed Central Google Scholar
Zimmerman, D. S. & Tan, H. L. Epidemiology and risk factors of sudden cardiac arrest. Curr. Opin. Crit. Care 27, 613–616 (2021).
Article PubMed Google Scholar
Andreka, P. & Frenneaux, M. P. Haemodynamics of cardiac arrest and resuscitation. Curr. Opin. Crit. Care 12, 198–203 (2006).
Article PubMed Google Scholar
Graham, R., McCoy, M. A. & Schultz, A. M. Strategies to improve cardiac arrest survival: A time to act. Strateg. Improv. Card. Arrest. Surviv. Time Act https://doi.org/10.17226/21723 (2015).
Article Google Scholar
Vaillancourt, C., Stiell, I. G. & Brien, S. Cardiac arrest care and emergency medical services in Canada. Can. J. Cardiol. 20, 1081–1090 (2004).
PubMed Google Scholar
Rea, T. D., Eisenberg, M. S., Sinibaldi, G. & White, R. D. Incidence of EMS-treated out-of-hospital cardiac arrest in the United States. Resuscitation 63, 17–24 (2004).
Article PubMed Google Scholar
Cobb, L. A., Fahrenbruch, C. E., Olsufka, M. & Copass, M. K. Changing incidence of out-of-hospital ventricular fibrillation, 1980–2000. JAMA 288, 3008–3013 (2002).
Article PubMed Google Scholar
Virani, S. S. et al. Heart disease and stroke statistics—2021 update. Circulation 143, E254–E743 (2021).
Article PubMed Google Scholar
Ferrari, M., Mottola, L. & Quaresima, V. Principles, techniques, and limitations of near infrared spectroscopy. Can. J. Appl. Physiol. 29, 463–487 (2004).
Article PubMed Google Scholar
Macnab, A. & Shadgan, B. Biomedical applications of wireless continuous wave near infrared spectroscopy. Biomed. Spectrosc. Imaging 1, 205–222 (2012).
Article Google Scholar
Hamaoka, T., McCully, K. K., Niwayama, M. & Chance, B. The use of muscle near-infrared spectroscopy in sport, health and medical sciences: Recent developments. Philos. Trans. R. Soc. A Math. Phys. Eng. Sci. 369, 4591–4604 (2011).
Article ADS CAS Google Scholar
Shadgan, B. et al. Optical assessment of spinal cord tissue oxygenation using a miniaturized near infrared spectroscopy sensor. 36, 3034–3043 (2019) https://home.liebertpub.com/neu.
Fischer, G. W. et al. Noninvasive cerebral oxygenation may predict outcome in patients undergoing aortic arch surgery. J. Thorac. Cardiovasc. Surg. 141, 815–821 (2011).
Article PubMed Google Scholar
Southern, W. M., Ryan, T. E., Reynolds, M. A. & Mccully, K. Reproducibility of near-infrared spectroscopy measurements of oxidative function and postexercise recovery kinetics in the medial gastrocnemius muscle. Appl. Physiol. Nutr. Metab. 39, 521–529 (2014).
Article CAS PubMed Google Scholar
Shah, A. & Shelley, K. H. Is pulse oximetry an essential tool or just another distraction? The role of the pulse oximeter in modern anesthesia care. J. Clin. Monit. Comput. 27, 235–242 (2013).
Article PubMed Google Scholar
Anderson, A. B., Zwerdling, R. G. & Dewitt, T. G. The clinical utility of pulse oximetry in the pediatric emergency department setting. Pediatr. Emerg. Care 7, 263–266 (1991).
Article CAS PubMed Google Scholar
Jubran, A. Pulse oximetry. Crit. Care 19, (2015).
Neff, T. A. Routine oximetry: A fifth vital sign?. Chest 94, 227 (1988).
Article CAS PubMed Google Scholar
Chan, E. D., Chan, M. M. & Chan, M. M. Pulse oximetry: Understanding its basic principles facilitates appreciation of its limitations. Respir. Med. 107, 789–799 (2013).
Article PubMed Google Scholar
Nitzan, M., Romem, A. & Koppel, R. Pulse oximetry: Fundamentals and technology update. Med. Devices (Auckl.) 7, 231–239 (2014).
PubMed Google Scholar
Allen, J. Photoplethysmography and its application in clinical physiological measurement. Physiol. Meas. 28, R1–R39 (2007).
Article ADS PubMed Google Scholar
Nickerson, B. G., Sarkisian, C. & Tremper, K. Bias and precision of pulse oximeters and arterial oximeters. Chest 93, 515–517 (1988).
Article CAS PubMed Google Scholar
Hakemi, A. & Bender, J. A. Understanding pulse oximetry, advantages, and limitations. Home Health. Care Manag. Pract. 17, 416–418 (2005).
Article Google Scholar
Fine, J. et al. Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring. Biosensors https://doi.org/10.3390/bios11040126 (2021).
Article PubMed PubMed Central Google Scholar
Ferrari, M., Mottola, L. & Quaresima, V. Principles, techniques, and limitations of near infrared spectroscopy. Can. J. Appl. Physiol. 29, 463–487. https://doi.org/10.1139/h04-031 (2011).
Article Google Scholar
Desanlis, J., Gordon, D., Calveyrac, C., Cottin, F. & Gernigon, M. Intra- and inter-day reliability of the NIRS portamon device after three induced muscle ischemias. Sensors 22, 5165 (2022).
Article ADS PubMed PubMed Central Google Scholar
Matcher, S. J., Elwell, C. E., Cooper, C. E., Cope, M. & Delpy, D. T. Performance comparison of several published tissue near-infrared spectroscopy algorithms. Anal. Biochem. 227, 54–68 (1995).
Article CAS PubMed Google Scholar
Ferraris, A., Jacquet-Lagrèze, M. & Fellahi, J. L. Four-wavelength near-infrared peripheral oximetry in cardiac surgery patients: A comparison between EQUANOX and O3. J. Clin. Monit. Comput. 32, 253–259 (2018).
Article PubMed Google Scholar
Watzman, H. M. et al. Arterial and venous contributions to near-infrared cerebral oximetry. Anesthesiology 93, 947–953 (2000).
Article CAS PubMed Google Scholar
Cheung, A. et al. Continuous optical monitoring of spinal cord oxygenation and hemodynamics during the first seven days post-injury in a porcine model of acute spinal cord injury. J. Neurotrauma 37, 2292–2301 (2020).
Article PubMed Google Scholar
Streijger, F. et al. Changes in pressure, hemodynamics, and metabolism within the spinal cord during the first 7 days after injury using a porcine model. J. Neurotrauma 34, 3336–3350 (2017).
Article PubMed Google Scholar
Kim, K. T. et al. Differences in morphometric measures of the uninjured porcine spinal cord and dural sac predict histological and behavioral outcomes after traumatic spinal cord injury. J. Neurotrauma 36, 3005–3017 (2019).
Article PubMed Google Scholar
Griffiths, J. R. & Robinson, S. P. The OxyLite: A fibre-optic oxygen sensor. Br. J. Radiol. 72, 627–630 (1999).
Article CAS PubMed Google Scholar
Leary, S. & Cartner, S. AVMA Guidelines for the Euthanasia of Animals: 2013 Edition Members of the Panel on Euthanasia. (2013).
Crellin, S. J. & Katz, K. D. Pentobarbital toxicity after self-administration of euthasol veterinary euthanasia medication. Case Rep. Emerg. Med. 2016, 1–4 (2016).
Article Google Scholar
Strøm, J., Häggmark, S., Reiz, S. & Sørensen, M. B. Cardiovascular effects of pentobarbital in pigs, and the lack of response to naloxone in pentobarbital induced circulatory failure. Acta Anaesthesiol. Scand. 31, 413–416 (1987).
Article PubMed Google Scholar
Urthaler, F., Krames, B. L. & James, T. N. Selective effects of pentobarbital on automaticity and conduction in the intact canine heart. Cardiovasc. Res. 8, 46–57 (1974).
Article CAS PubMed Google Scholar
Ibrahim, A. W. et al. Cerebral oximetry as a real-time monitoring tool to assess quality of in-hospital cardiopulmonary resuscitation and post cardiac arrest care. J. Am. Heart Assoc. 4, e001859 (2015).
Article PubMed PubMed Central Google Scholar
Asim, K. et al. Near infrared spectrophotometry (cerebral oximetry) in predicting the return of spontaneous circulation in out-of-hospital cardiac arrest. Am. J. Emerg. Med. 32, 14–17 (2014).
Article PubMed Google Scholar
Goto, Y., Funada, A. & Goto, Y. Relationship between the duration of cardiopulmonary resuscitation and favorable neurological outcomes after out-of-hospital cardiac arrest: A prospective, nationwide, population-based cohort study. J. Am. Heart Assoc. 5, (2016).
Takegawa, R., Hayashida, K., Choudhary, R., Rolston, D. M. & Becker, L. B. Brain monitoring using near-infrared spectroscopy to predict outcome after cardiac arrest: A novel phenotype in a rat model of cardiac arrest. J. Intensive Care 9, 1–3 (2021).
Article Google Scholar
Drennan, I. R., Gilgan, J., Goncharenko, K. & Lin, S. Use of near-infrared spectroscopy by paramedics during out-of-hospital cardiac arrest: A feasibility study. CJC Open 1, 256–260 (2019).
Article PubMed PubMed Central Google Scholar
Callaway, C. W. Cerebral oximetry and cardiopulmonary resuscitation. J. Am. Heart Assoc. 4, 1–2 (2015).
Article Google Scholar
Cournoyer, A. et al. Near-infrared spectroscopy monitoring during cardiac arrest: A systematic review and meta-analysis. Acad. Emerg. Med. 23, 851–862 (2016).
Article PubMed Google Scholar
Ahn, A., Nasir, A., Malik, H., D’Orazi, F. & Parnia, S. A pilot study examining the role of regional cerebral oxygen saturation monitoring as a marker of return of spontaneous circulation in shockable (VF/VT) and non-shockable (PEA/Asystole) causes of cardiac arrest. Resuscitation 84, 1713–1716 (2013).
Article PubMed Google Scholar
Mick, N. W. & Williams, R. J. Pediatric cardiac arrest resuscitation. Emerg. Med. Clin. N. Am. 38, 819–839 (2020).
Article Google Scholar
OxyLiteTM dissolved oxygen and temperature monitors|Oxford Optronix. https://www.oxford-optronix.com/oxylite-oxygen-monitors.
Sanni, A. A. & McCully, K. K. Interpretation of near-infrared spectroscopy (NIRS) signals in skeletal muscle. J. Funct. Morphol. Kinesiol. 4, (2019).
Cheung, A., Tu, L., Macnab, A., Kwon, B. K. & Shadgan, B. Detection of hypoxia by near-infrared spectroscopy and pulse oximetry: A comparative study. J. Biomed. Opt. 27, (2022).
Aron, J. H. et al. Cerebral oxygen desaturation after cardiopulmonary bypass in a patient with Raynaud's phenomenon detected by near-infrared cerebral oximetry. Anesth. Analg. 104, 1034–1036 (2007).
Article PubMed Google Scholar
Rodríguez, A. et al. Regional oxygen saturation index (rSO2) in brachioradialis and deltoid muscle. Correlation and prognosis in patients with respiratory sepsis. Med. Intensiva 39, 68–75 (2015).
Article PubMed Google Scholar
Rodriguez, A. et al. Mortality and regional oxygen saturation index in septic shock patients: A pilot study. J. Trauma Inj. Infect. Crit. Care 70, 1145–1152 (2011).
CAS Google Scholar
Crookes, B. A. et al. Can near-infrared spectroscopy identify the severity of shock in trauma patients?. J. Trauma 58, 806–816 (2005).
Article PubMed Google Scholar
Aleman, M., Williams, D. C., Guedes, A. & Madigan, J. E. Cerebral and brainstem electrophysiologic activity during euthanasia with pentobarbital sodium in horses. J. Vet. Intern. Med. 29, 663–672 (2015).
Article CAS PubMed PubMed Central Google Scholar
Donaldson, A. E. & Lamont, I. L. Biochemistry changes that occur after death: Potential markers for determining post-mortem interval. PLoS ONE 8, 82011 (2013).
Article ADS Google Scholar
Barksdale, K. A. et al. Mitochondrial viability in mouse and human postmortem brain. FASEB J. 24, 3590–3599 (2010).
Article CAS PubMed PubMed Central Google Scholar
Olgin, J. E. et al. Wearable cardioverter-defibrillator after myocardial infarction. N. Engl. J. Med. 379, 1205–1215 (2018).
Article PubMed PubMed Central Google Scholar
Zishiri, E. T. et al. Early risk of mortality after coronary artery revascularization in patients with left ventricular dysfunction and potential role of the wearable cardioverter defibrillator. Circ. Arrhythm. Electrophysiol. 6, 117 (2013).
Article PubMed Google Scholar
Lee, Y., Shin, H., Choi, H. J. & Kim, C. Can pulse check by the photoplethysmography sensor on a smart watch replace carotid artery palpation during cardiopulmonary resuscitation in cardiac arrest patients? A prospective observational diagnostic accuracy study. BMJ Open 9, e023627 (2019).
Article PubMed PubMed Central Google Scholar
Castaneda, D., Esparza, A., Ghamari, M., Soltanpur, C. & Nazeran, H. A review on wearable photoplethysmography sensors and their potential future applications in health care. Int. J. Biosens. Bioelectron. 4, 195 (2018).
PubMed PubMed Central Google Scholar
Blok, S., Piek, M. A., Tulevski, I. I., Somsen, G. A. & Winter, M. M. The accuracy of heartbeat detection using photoplethysmography technology in cardiac patients. J. Electrocardiol. 67, 148–157 (2021).
Article CAS PubMed Google Scholar
Pereira, T. et al. Photoplethysmography based atrial fibrillation detection: A review. NPJ Digit. Med. 3, 1–12 (2020).
Article Google Scholar
Chorin, E. et al. Continuous heart rate monitoring for automatic detection of life-threatening arrhythmias with novel bio-sensing technology. Front. Cardiovasc. Med. 1, 748 (2021).
Google Scholar
Castaneda, D., Esparza, A., Ghamari, M., Soltanpur, C. & Nazeran, H. A review on wearable photoplethysmography sensors and their potential future applications in health care. Int. J. Biosens. Bioelectron. 4, (2018).
DeMeulenaere, S. Pulse oximetry: Uses and limitations. J. Nurse Pract. 3, 312–317 (2007).
Article Google Scholar
Spittal, M. J. Evaluation of pulse oximetry during cardiopulmonary resuscitation. Anaesthesia 48, 701–703 (1993).
Article CAS PubMed Google Scholar
Chan, J., Rea, T., Gollakota, S. & Sunshine, J. E. Contactless cardiac arrest detection using smart devices. NPJ Digit. Med. 2, (2019).
Eisenberg, M. S. Incidence and significance of gasping or agonal respirations in cardiac arrest patients. Curr. Opin. Crit. Care 12, 204–206 (2006).
Article PubMed Google Scholar
Download references
The authors gratefully acknowledge the staff at the UBC Kwon Research Lab and UBC Center for Comparative Medicine, who supported these complex in vivo experiments and provided care for the animals involved in the studies. This study was supported by a Translational Research Award from the U.S. Department of Defense, Spinal Cord Injury Research Program (SCIRP). Mahsa Khalili holds a Trainee Award, Brian Grunau holds a Health Professional-Investigator Award, and Babak Shadgan holds a Scholar Award from the Michael Smith Health Research BC.
This study was supported by a Translational Research Award from the U.S. Department of Defense, Spinal Cord Injury Research Program (SCIRP), SC130007, the Michael Smith Foundation for Health Research (MSFHR), Award No. AWD-00482, the Defense Advanced Research Projects Agency (DARPA), Contract No. N660012024046 and MITACS Award No. IT25703. The views, opinions and/or findings expressed are those of the author and should not be interpreted as representing the official views or policies of the Department of Defense or the U.S. Government.
International Collaboration on Repair Discoveries (ICORD), University of British Columbia, 818 West 10th Avenue, Vancouver, BC, V5Z 1M9, Canada
Katharina Raschdorf, Arman Mohseni, Amanda Cheung, Kitty So, Neda Manouchehri & Babak Shadgan
Department of Neuroscience, University of British Columbia, 2215 Wesbrook Mall, Vancouver, BC, V6T 1Z3, Canada
Katharina Raschdorf & Babak Shadgan
School of Biomedical Engineering (SBME), University of British Columbia, 2222 Health Sciences Mall, Vancouver, BC, V6T 1Z4, Canada
Kaavya Hogle, Saud Lingawi, Calvin Kuo & Babak Shadgan
Department of Emergency Medicine, University of British Columbia and St. Paul's Hospital, Vancouver, BC, Canada
Mahsa Khalili, Brian Grunau & Jim Christenson
Department of Orthopaedics, University of British Columbia, 2775 Laurel Street, Vancouver, BC, V5Z 1M9, Canada
Babak Shadgan
You can also search for this author in PubMed Google Scholar
You can also search for this author in PubMed Google Scholar
You can also search for this author in PubMed Google Scholar
You can also search for this author in PubMed Google Scholar
You can also search for this author in PubMed Google Scholar
You can also search for this author in PubMed Google Scholar
You can also search for this author in PubMed Google Scholar
You can also search for this author in PubMed Google Scholar
You can also search for this author in PubMed Google Scholar
You can also search for this author in PubMed Google Scholar
You can also search for this author in PubMed Google Scholar
You can also search for this author in PubMed Google Scholar
Conceptualization, B.S. and K.R.; methodology, K.R.; formal analysis, K.R.; analysis validation, B.G., K.C., J.C. and B.S.; data collection and curation, B.S., K.R., K.S., A.C., N.M.; writing (original draft preparation), K.R. and A.M.; writing (review and editing), B.S., C.K., B.G., J.C., M.K., and S.L.; visualization, K.R. All authors have read and agreed to the published version of the manuscript.
Correspondence to Babak Shadgan.
The authors declare no competing interests.
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
Reprints and Permissions
Raschdorf, K., Mohseni, A., Hogle, K. et al. Evaluation of transcutaneous near-infrared spectroscopy for early detection of cardiac arrest in an animal model. Sci Rep 13, 4537 (2023). https://doi.org/10.1038/s41598-023-31637-1
Download citation
Received: 31 July 2022
Accepted: 15 March 2023
Published: 20 March 2023
DOI: https://doi.org/10.1038/s41598-023-31637-1
Anyone you share the following link with will be able to read this content:
Sorry, a shareable link is not currently available for this article.
Provided by the Springer Nature SharedIt content-sharing initiative
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.